Elemental (CHN) Analyzer LCHN-A10 is used to determine Carbon, Hydrogen and Nitrogen in macro amount of samples. Features dual stage furnace with pure oxygen flow for complete combustion, multiple detector, blended gas analysis, IR CO2 monitor detector, stackable auto loader etc. to ensure efficient performance and obtain reliable results.
Specifications
Gas required | Helium: 99.99% 0.25 ± 0.01 mPa Oxygen: 99.5% 0.25 ± 0.01 mPa Nitrogen or Composed air: 0.25 ± 0.01 mPa |
Measurement range | Carbon: 0.02 to 150 mg Hydrogen: 0.1 to 12 mg Nitrogen: 0.04 to 50 mg |
Consumption | Helium 200 ml/ min |
Sample mass | Up to 1000 mg, depending on the sample matrix |
Analysis time | 4 to 6 mins, depending on sample combustion condition |
Autosampler | Stackable auto loader, up to 140 samples by 4 layers |
Accuracy | Carbon (Cad): 0.45% Hydrogen (Had): 0.10% Nitrogen (Nad): 0.05% |
Furnace type | Resistance furnace (main furnace and furnace rear) |
Maximum temperature | 1050°C |
Dimension | 690 × 750 × 720 mm |
Weight | 110 kg |
Power | 5.5 kW |
Power supply | AC 220 ± 10 %V , 50/60 Hz |
Features
Dual-stage furnace with complete catalytic combustion
Customized multiple detectors for different elements
Sample is analyzed in minimum time
5.0 ml Blended gas analysis to reduce the consumption of helium and copper
IR CO2 and IR H2O detector are used
Complete combustion ensured by IR CO2 monitor detector
Stackable auto loader , 25/35 samples in each layer
Can analyze macro weighed samples of liquids, solids and heterogeneous matrix
Windows based operating software to control and view parameters on PC
Complies with CE standards
Applications
It is used for analysis of fuels, biomass, food, oil, coal, waste, plastic, solvents, fertilizers etc. samples in different laboratories and industries.
Request Quote