ICP spectrometer LICP-A10

ICP spectrometer LICP-A10 detects trace elements using argon plasma where the atomized liquid sample is injected. With an analyzing rate of 26 elements per minute results can be achieved with lower levels of detection and measurement of trace and ultra-trace elements. Elements can be traced from liquid and solid samples. Detection of liquid flow from the spray chamber by optical drain sensor offers exceptional sensitivity and accuracy.

Specifications Features Applications Element Detection Limits (Unit - µg / ml) Get Quote View Catalog

Specifications

Wavelength range ( 3600 lines / mm ) 180 to 500 nm
Wavelength range ( 2400 lines / mm ) 180 to 800 nm
Temperature 20 ~ 28 °C
Elements per minute 26 elements
Humidity 70%
Focal length 1000 mm
Standard deviation RSD ≤ 1.0 %
Correlation coefficient ≥ 0.9998
Gas source Argon ( 99.99 % )
Plasma gas flow meter 100 ~ 1000 L / h
Auxillary gas flow meter 10 ~ 100 L / h
Carrier gas flow meter 10 ~ 100 L / h
Cooling water flow rate 5 liters per min
Cooling water temperature 21 ~ 26 °C
Water outlet diameter Ø 10 mm
Line dispersion rate 0.26 nm
Resolving power ≤ 0.008 nm
Incident slit 20 µm
Co-axial sprayer diameter 6 mm
Coil diameter 25 mm ( 3 turn )
Double room fog diameter 34 mm
Quartz tube torch diameter 20 mm
Mirror dimensions 78 x 105 x 16 mm

Features

Scanning range variance ( 190 nm to 500 nm / 190 nm to 800 nm )

Rapid analysis – 26 elements per minute

Temperature range – 20 to 28 °C

Optical pathway – Czerny-turners

Low detection limit

Detection of multi-elements including non-metallic

Single and multi-element measurement method

Indication for wavelength error and repeatability

Range of measurement – common to trace

Applications

Used in clinical analysis for detection of metals in biological fluids , environmental analysis for trace element detection in water , soil , plants , pharmaceuticals for traces of catalysts and poison metals like Cd , Pb , industries for noble metal detection

Element Detection Limits (Unit - µg / ml)

Element Be Mg Ca Sr Ba Sc Y Eu Yb
Wavelength ( nm ) 313.042 279.553 393.366 407.771 455.403 335.373 371.03 381.967 369.419
Detection limit < 1.0 < 1.0 < 1.0 < 1.0 < 1.0 < 1.0 < 1.0 < 1.0 < 1.0
Element Lu La Gd Dy Ho Er Tm Zn Cd
Wavelength ( nm ) 261.542 333.749 342.247 353.17 345.6 337.271 313.126 213.856 226.502
Detection limit < 1.0 < 3.0 < 3.0 < 3.0 < 3.0 < 3.0 < 3.0 < 3.0 < 3.0
Element Fe Ti Cu Mn V B Ni Cr Co
Wavelength ( nm ) 239.562 344.941 324.754 257.61 309.311 249.773 221.647 205.552 238.204
Detection limit < 3.0 < 3.0 < 3.0 < 3.0 < 3.0 < 5.0 < 5.0 < 5.0 < 5.0
Element Zr Mo Ag Au Si Al Ir P Hg
Wavelength ( nm ) 343.823 202.03 328.068 242.795 251.611 396.152 224.268 213.618 194.227
Detection limit < 5.0 < 5.0 < 5.0 < 5.0 < 5.0 < 10.0 < 10.0 < 10.0 < 10.0

Request Quote

ICP Spectrometer

ICP Spectrometer LICP-B10

Monochromator Specifications
Optical type Czerny turner
Resolution ≤ 0.015nm (3600 line grating) ≤ 0.030nm (2400 line grating)
Focal length 1000 mm
Grating specifications huge holographic grating with 3600 L/mil or 2400 L/mil and 80 mm × 100 mm of ruling area
Wavelength range 195 to 500 nm for 3600 line grating 195 to 800 nm for 2400 line grating
Solid State Power Specifications
Frequency 27.12 MHz Frequency stability: < 0.05%
Spray chamber Scott double pass spray chamber
Output power 800 W to 1600 W, adjustable with power efficiency more than 65%
Output power stability ≤ 0.05%
Induction coil 25 mm × 3 ID (ID-internal diameter), equipped with three concentric quartz torch tubes 35 mm ED (ED-external diameter)
Technical Specifications
Suitable sample content range Liquid sample: 0.01 ppm to several thousand ppm Solid or power sample: 0.001% to 70%
Repeatability short-term stability- RSD ≤ 1.5% Long-term stability: RSD ≤ 2%
Test speed 5 to 8 elements / min
Limits of detection (LOD, µg/L) for most elements 1ppb to 10ppb
Control Circuit
Photomultiplier tube specification R212 / R928
Negative high voltage -50 to -1000 V
Circuit measuring range (10-12 to 10 -4) A
Signal acquisition VF conversion
Working Environment
Temperature for storage and transportation 15°C to 25°C
Humidity for storage and transportation ≤ 70%
Power Stability 220 ± 10V; 50 to 60 V
Humidity ≤ 70%
Temperature 15°C to 30°C
Other Details
Packaging Dimension and Gross Weight (4 case) 1st case: 188 × 111 × 105 / 348 kg 2nd case: 60 × 60 × 55 / 25 kg 3rd case: 59 × 67 × 97 / 80 kg 4th case: 38.5 × 38.5 × 63 / 47 kg

ICP Spectrometer LICP-A11

Wavelength range 180 to 900 nm
Elements per minute 15 elements
Temperature 29 °C
Incident slit 25 µm
Standard deviation RSD ≤ 2 %
Correlation coefficient ≥ 0.9995 %
Gas source Argon ( 99.99 % )
Plasma gas flow meter 100 – 1000 L / h ( 1.6 - 16 L / min )
Auxillary gas flow meter 10 – 100 L / h ( 0.16 – 1.66 L / min )
Carrier gas flow meter 0 ~ 5 L / min
Cooling water flow rate 5 L / min
Cooling water temperature 20 ~ 26 °C
Water outlet diameter 10 mm
Line dispersion rate 0.26 nm
Resolving power ≤ 0.007 nm
Co-axial sprayer diameter 6 mm
Coil diameter 25 mm
Double room fog diameter 34 mm
Quartz tube torch diameter 20 mm
Mirror dimensions 80 x 110 x 16 mm
ICP Spectrometer LICP-A12
Wavelength range 195 to 800 mm
Elements per minute 10 elements
Temperature range 32 °C
Incident slit 20 µm
Standard deviation RSD ≤ 2 %
Correlation coefficient ≥ 0.9995 %
Gas source Argon ( 99.99 % )
Plasma gas flow meter 100 ~ 1000 L / h ( 1.6 - 16 L / min )
Auxillary gas flow meter 10 ~ 100 L / h ( 0.16 – 1.66 L / min )
Carrier gas flow meter 0 ~ 5 L / min
Cooling water flow rate 5 L / min
Cooling water temperature 20 ~ 26 °C
Water outlet diameter 10 mm
Line dispersion rate 0.26 nm
Resolving power ≤ 0.008 nm
Co-axial sprayer diameter 6 mm
Coil diameter 25 mm ( 3 turn )
Double room fog diameter 34 mm
Quartz tube torch diameter 20 mm
Mirror dimensions 78 x 105 x 16 mm

XRF Spectrometer

XRF-Spectrometer LXRF-A10

Measureable elements S to U
Detection limit 1 ppm
Temperature 15 ~ 30 °C
Elemental content 1 ppm to 99 %
Repeatability 0.001
Stability 0.001
Power supply AC 220 V ± 5 V
Dimensions 550 x 410 x 320 mm
Weight 45 kg

Handheld XRF Spectrometer LXRF-B10

Analytical Method Energy dispersive X-ray fluorescence analytical Method
Elements Measuring Atomic number from 12 to 92 [elements from magnesium (Mg) to uranium (U)] can be measured
Range Simultaneous detector
Content measurement range ppm to 100%
Detector SDD detector, Simultaneous analysis 40 elements
Detector resolution Minimum can reach 128 eV
Detection limit Minimum detection limits at 1 to 500 ppm
Detection Objective Solid, liquid, powder Detection Objective Solid, liquid, powder
The detection time 1 to 60 seconds (a second report results)
Excitation source 50 kV/200 uA- silver target end window integrated miniature X ray tube and high voltage power supply
Collimator and filter Collimator diameters are 4.0 mm and 2.0 mm, 6 kinds of filters with automatic switching functions
Camera system Built-in 500 W pixels high definition camera, to observe testing position at any time
Display screen Brand new 5 inch transflective LCD touch screen, the resolution is 1080×720
Built-in system GPS, WIFI, Bluetooth
Power Supply Rechargeable lithium battery, standard is 9000 mAh, sustainable work up to 12 hours
Ambient temperature -20 to 50°C
Relative Humidity ≤ 90%
Data transmission Digital multi-channel technology, SPI data transmission, quick analysis, the high count rate; waterproof mini USB, and can be connected with a desktop computer
Warning-Signalling System Green light means power on, red flushing means testing and yellow flushing means the problems
Dimension (L×W×H) 244 × 90 × 330 mm
Weight 1.7 kg
Handheld XRF Spectrometer LXRF-B11
Analytical Method Energy dispersive X-ray fluorescence analytical Method
Elements Measuring Atomic number from 12 to 92 [elements from magnesium (Mg) to uranium (U)] can be measured
Range Simultaneous detector
Content measurement range ppm to 99.99%
Detector SDD detector or fast SDD detector (optional)
Detector resolution Minimum can reach 128 eV
Detection limit Minimum detection limits at 1 to 500 ppm
Detection Objective Solid, liquid, powder Solid, liquid, powder
The detection time 1 to 60 seconds (a second report results)
Excitation source 50 kV/200 uA- silver target end window integrated miniature X ray tube and high voltage power supply
Collimator and filter Collimator diameters are 4.0 mm and 2.0 mm, 6 kinds of filters with automatic switching functions
Camera system Built-in 500 W pixels high definition camera, to observe testing position at any time
Display screen Brand new 5 inch transflective LCD touch screen, the resolution is 1080×720
Built-in system GPS, WIFI, Bluetooth
Power Supply Rechargeable lithium battery, standard is 27000 mAh, sustainable work up to 12 hours
Ambient temperature -20 to 50°C
Relative Humidity ≤ 90%
Data transmission Digital multi-channel technology, SPI data transmission, quick analysis, the high count rate; waterproof mini USB, and can be connected with a desktop computer
Warning-Signalling System Green light means power on, red flushing means testing and yellow flushing means the problems
Dimension (L×W×H) 244 × 90 × 330 mm
Weight 1.7 kg