Ion chromatograph system LICS-A22 with cation and anion dual channel design, independently operates and simultaneously detects both channels without disturbing each other. Temperature controlled bipolar conductivity detector, improves the detection performance and stability of the instrument. System is comprised of an auto sampler, anion and cation suppressor.
Used to detect small molecules (in the form of anion, cation) in water testing & processing plants, food, chemical and dye industry, mining, metallurgy, and other fields.
| Ion Chromatographic Pump | {"Type":"High pressure and low pulse double plunger tandem advection pump","Maximum pressure":"35 MPa (PEEK)","Pressure display accuracy":"\u2264 0.1 MPa","Flow stability":"(0.2 to 0.5) ml\/ min \u2264 3 % \n(0.5 to 1.0) ml\/ min \u2264 2 % \n> 1.0 ml\/ min 2 % ","Flow range":"0.001 to 9.999 ml\/ min","Pressure pulse":"\u2264 0.5%"} |
| Conductivity detector | {"Type":"Temperature Control and Bipolar Conductivity Detector","Detection mode":"Bipolar conductivity detection","Cell volume":"\u2264 0.8 \u00b5L","Detection range":"0 to 35000 \u00b5S\/ cm","Detection resolution":"\u2264 0.0020 nS\/ cm","Output voltage":"- 6000 to +6000 mV (adjustable)","Baseline noise":"\u2264 0.05 % F.S.","Baseline drift":"\u2264 3 % FS\/ 30 min","Operating temperature range":"Room temperature +5 to 60\u00b0C (41 to 140\u00b0F)","Controlling temperature accuracy":"\u00b1 0.01\u00b0C","Maximum pressure":"10 MPa","Linear range":"\u2265 103","Instruments linearity":"\u2265 0.999","Quantitative repeatability":"\u2264 1.0 %","Qualitative repeatability":"\u2264 1.0 %","Minimum detectable concentration":"Cl \u2264 0.0005 ug\/ ml \nBrO3 \u2264 0.001 ug\/ ml \nLi+ \u2264 0.001 ug\/ ml"} |
| Flow system | {"Plastic flow path":"Made of PEEK materials","Six way Valve":"PEEK material, 5000 psi pressure, independent automatic collection and flow function"} |
| Degassing Device | {"Inner volume":"300 \u00b5l","Degassing volume":"7.5 ml","Maximum flow rate":"10 ml\/ min","Degassing efficiency":"Degassing 90 % at 1ml\/ min","Organic tolerance":"1"} |
| Ion Chromatographic Pump | {"Type":"High pressure and low pulse double plunger tandem advection pump","Maximum pressure":"35 MPa (PEEK)","Pressure display accuracy":"\u2264 0.1 MPa","Flow stability":"(0.2 to 0.5) ml\/ min \u2264 3 % \n(0.5 to 1.0) ml\/ min \u2264 2 % \n> 1.0 ml\/ min 2 %","Flow range":"0.001 to 9.999 ml\/ min","Pressure pulse":"\u2264 0.5%"} |
| Conductivity detector | {"Type":"Temperature Control and Bipolar Conductivity Detector","Detection mode":"Bipolar conductivity detection","Cell volume":"\u2264 0.8 \u00b5L","Detection range":"0 to 45000 \u00b5S\/ cm","Detection resolution":"\u2264 0.0020 nS\/ cm","Output voltage":"- 6000 to +6000 mV (adjustable)","Baseline noise":"\u2264 0.05 % F.S.","Baseline drift":"\u2264 3 % FS\/ 30 min","Operating temperature range":"Room temperature +5 to 60\u00b0C (41 to 140\u00b0F)","Controlling temperature accuracy":"\u00b1 0.01\u00b0C","Maximum pressure":"10 MPa","Linear range":"\u2265 103","Instruments linearity":"\u2265 0.999","Quantitative repeatability":"\u2264 1.0 %","Qualitative repeatability":"\u2264 1.0 %","Minimum detectable concentration":"Cl \u2264 0.0005 ug\/ ml \nBrO3 \u2264 0.001 ug\/ ml \nLi+ \u2264 0.001 ug\/ ml"} |
| Flow system | {"Plastic flow path":"Made of PEEK materials","Six way Valve":"PEEK material, 5000 psi pressure, independent automatic collection and flow function"} |
| Power supply | 150 W |
| Oven | {"Maximum capacity":"22 L","Temperature range":"RT + 5 \u2103 to 400 \u2103","Temperature accuracy":"\u00b1 0.1 \u2103","Temperature program":"9-phase\/ 10 platform","Program total time":"9999.9 min","Maximum temperature heating rate":"0.1 to 60 \u2103\/ min","Maximum cooling rate":"\u02c2 10 mins (250 to 50 \u2103)"} |
| Sample injector | {"Temperature range":"RT+ 7 \u2103 to 420 \u2103","Temperature control":"Automatic","Injector type":"Column or split","Maximum sample inlet":"3 pcs","Carrier gas flow type":"Constant pressure","Pre column pressure range":"0 to 400 kPa","Pre column pressure accuracy":"0.1 kPa","Mobile phase flow range":"H2: 0 to 200 ml\/min, N2: 0 to 150 ml\/min"} |
| Detector | {"Temperature range":"RT + 7 \u2103 to 420 \u2103","Maximum installation":"Dual detector","Flame ignition mode":"Automatic","Flame Ionization Detector (FID)":"Logarithmic amplifier \nHigh voltage switch control \nBaseline signal display \nIgnition coil control \nDetection limit: \u22643 \u00d710 to 12 g\/s \n(Sample: n-hexadecane) \nBaseline noise: \u2264 5 \u00d7 10 - 14 <\/sup> A \nBaseline draft: \u2264 6 \u00d7 10 -13 <\/sup> A \nDynamic range: 10 7 <\/sup> \nRSD: \u2264 3 %"} |
| Power supply | AC 220 V ± 22 V, 50 Hz ± 0.5 Hz, 3000 W |
| Gas Chromatography Specifications | {"Model No":"GC-87","Flow rate":"1 ~ 1.5 ml \/ min","Maximum flow rate":"10 ml \/ min","Inlet temperature":"450 \u00b0C","Pressure range":"0 ~ 100 psi","Heating rate":"Upto 120 \u00b0C \/ min","Room temperature":"4 \u00b0C ~ 450 \u00b0C","Pressure control mode":"Electronic pressure control ( Supports CV and CC )","Split mode":"Split \/ split less","Split ratio":"1000 : 1","Temperature programming":"7 stages \/ 8 platforms","Auto sampler":"Optional"} |
| Mass Spectrometry Specifications | {"Model No":"MS-89","Mass range":"1.5 ~ 1000 amu","Ion \u2013 source temperature":"100 \u00b0C ~ 350 \u00b0C","GC-MS interface temperature":"450 \u00b0C","Scan rate":"Upto 10000 amu\/s","Stability":"\u00b1 0.10 amu \/ 48 hrs.","Filament emission current":"0 ~ 350 \u00b5A","EI source ionization energy":"5 eV ~ 250 eV","Resolution":"Unit resolution","Detector":"High energy dynode electron multiplier","Sensitivity":"Full scan ( S \/ N is \u2265 30 : 1 )","Vacuum":"Turbo molecular pump ( 67 L\/s )"} |
| Chromatographic Pump | {"Type":"High-pressure and low \u2013pulse double plunger tandem advection pump","Maximum Pressure":"35 MPa (PEEK)","Flow Range":"0.001 to 9.999 mL\/min","Flow accuracy":"\u00b1 0.5 %","Flow Repeatability":"RSD \u2264 0.1%","Flow stability":"(0.2 to 0.5) mL\/min, \u2264 3 % \n(0.5 to 1.0) mL\/min, \u2264 2 % \n>1.0 mL\/min, \u2264 2 %"} |
| Built-in Heat conduction type column heater | {"Temperature range":"Room temperature +5\u00b0C to 60\u00b0C (41 to 140\u00b0F)","Allowable deviation":"\u00b1 1\u00b0C","Temperature stability":"\u2264 0.5\u00b0C\/h"} |
| Electromagnetic injection valve | {"Maximum pressure":"35 MPa","Control mode":"By stepper motor","Power supply":"24 V (DC)"} |
| Digital and temperature control detection system | {"Type":"Temperature control and bipolar conductivity detector","Detection type":"Bipolar conductivity detection","Cell volume":"\u2264 0.8 \u00b5L","Measurement range":"0 to 45000 \u00b5 S\/cm (adjustable)","Detection resolution":"\u2264 0.0020 nS\/cm","Output voltage":"-6000 to 6000 mV (adjustable)","Temperature range":"Room temperature +5\u00b0C to 60\u00b0C (41 to 140\u00b0F)","Temperature accuracy":"\u00b1 0.01\u00b0C","Maximum pressure":"10 MPa","Baseline drift":"\u2264 20 % FS\/30 min","Baseline noise":"\u22640.5 % FS","Instrument linearity":"\u2265 0.999","Quantitative Repeatability":"\u22640.5 %","Qualitative Repeatability":"\u22642 %","Minimum detectable concentration":"Cl-\u22640.005 \u00b5g\/mL \nLi+ \u22640.005 \u00b5g\/mL"} |